| KW | Description |
| mc.Status | Mirror Cooling State or Mode: Off, Manual, Auto
Auto has 3 states: Auto_ON, auto_DEROST, Auto_OFF
|
| mc.State.Tm | Time spend in current mc.status or mode. |
| mc.SetPt | MC Setpt is target temperature for the mirror |
| mc.auto.SetPt | If ON, the mc.SetPt is set using data from MKWC
age is how old the auto set point value is. |
| mc.Defrost.Cnt | Counter to indicated Number of Defrost cycles
Auto will switch into OFF state after too many defrost cycles. |
| mc.Warm.Air.Cnt | Counter to indicated 'Air too warm' cycles
Auto will switch into OFF state after too many tries. |
| MC.TAO | Mirror Cooling Temperature Air Out Deg C |
| MC.TG | Mirror Cooling Temperature Glycol (In), Deg C |
| MC.TGO | Mirror Cooling Temperature Glycol Out, Deg C |
| MC.Act | Mirror Cooling Actuator Volts. 0v=close, 10v=open |
| MC.FAN | Mirror Cooling Heater, OFF or ON |
| MC.HEAT | Mirror Cooling Fan, OFF or ON |
| MC.Pressure.Err | Mirror Cooling Pressure Differental sensor. ON for error. |
| HE1 | IQUP HE1 humidity data |
| HD.avg | IQUP Humidity Dome. Average of HD1 thur HD3. |
| HM.est | Humidity in Mirror Cell. Caculated from DewPt |
| DewPt | IQUP DewPt Sensor in Mirror Cell, Deg C. |
| TE1 | IQUP Temperature Environment# 1, Deg C |
| TD.h | IQUP Dome High Dome, average of (TD1, TD2). Deg C |
| TD.m | IQUP Temperature Dome Mid, average of (TD3,TD4,TD5). Deg C |
| TD.l | IQUP Temperature Dome Low, average of (TD6, TD7). Deg C |
| TMSavg | IQUP Temperature Mirror Cell, average of (TMS1, TMS2). Deg C |
| TM.avg | IQUP Temperature Mirror, average of (TM1 to TM6). Deg C |
| TF.avg | IQUP Temperature Floor, average of (TF1 to TF4). Deg C |